4 display lock, 5 control panel 2 - electric coarse feed, Display lock – Leica Biosystems CM1860/CM1860 UV User Manual

Page 34: Control panel 2 - electric coarse feed, Operating the instrument

Advertising
background image

34

Instructions for Use V1.4 – 10/2012

Advancing the specimen toward the knife
• Starts fast or slow feed toward the knife. The LED (2) starts flashing as

long as the specimen head is in motion.

The LED (2) lights up when the forward end position has been reached.

This is a safety feature designed to protect both the specimen and

the knife.

• To feed the specimen, press or hold the appropriate SLOW or RAPID

button.

6.5 Control panel 2 - Electric coarse feed

Move the specimen away from the knife
• Starts fast retraction to the rear end position.

The LED (1) flashes while the specimen head is in motion.

The LED (1) lights up when the rear end position has been reached.

• The return movement can be stopped by pressing one of the coarse feed

buttons.

• Starts slow retraction to the rear end position.

The motion continues as long as the button is held.

6.

Operating the Instrument

1

2

slow

rapid

Fig. 23

slow

rapid

Fig. 22

6.4 Display lock

After the display is locked using the key button (press and hold for 5 sec.),

set values can no longer be changed.
Only the UV disinfection can still be enabled/disabled by pressing the UV key.
Push the key button once more for 5 seconds to unlock the display.
When the display is locked, the LEDs between the hour and minute indication

on the time panel are turned off.