Leica CM1860/CM1860 UV – Cryostat

59

8. Troubleshooting

Problem

Cause

Remedy

Condensation on anti-roll plate and

knife during cleaning

- Brush, forceps and/or cloth are

too warm.

- Use components and clean-

ing materials only after cool-

ing them.

Store all tools on storage

shelf in the cryochamber.

Anti-roll plate damaged after ad-

justment

- Plate too high above the cutting

edge. The adjustment was car-

ried out in the direction of the

knife.

- Replace anti-roll plate; then

lift it during adjustment.

- Handle anti-roll plate more

carefully.

Thick-thin sections

- Temperature incorrect for the

tissue cut.

- Knife profile inappropriate for the

specimen cut.

- Ice buildup in the knife back.

- Handwheel speed not uniform or

turned at incorrect speed.

- Knife not clamped tightly enough.

- Specimen holder not clamped

tightly.

- Cryocompound applied to cold

specimen disc; specimen de-

tached from the disc after freez-

ing.

- Blunt cutting edge.

- Incorrect clearance angle.

- Dried specimen.

- Select correct temperature.

- Wait until the correct tem-

perature is reached.

- Use knife with different pro-

file (c or d).

- Remove ice.

- Adapt speed.

- Check knife/blade clamping.

- Check specimen clamping.

- Apply cryocompound on

warm disc; mount specimen

and freeze.

- Use different part of the cut-

ting edge or replace.

- Slightly increase clearance

angle.

- Prepare new specimen.