2.3 Safety features on the instrument

The Leica ASP200 tissue processor is equipped with numerous safety func-

tions and with sophisticated software control mechanisms. These ensure

that the specimens remain undamaged in the event of a power failure or

other malfunction during the infiltration process and that the infiltration is

completed successfully.

Overpressure protection

• When power is off the air pump and air valves default to a safe condition

(retort vented, no pressure generation).

• If the pump is not stopped at the correct moment during the pressuriza-

tion, the power supply is interrupted by a separate electronic circuit.

• In addition, there is a safety relief valve that vents all excess air pump

output to atmosphere.

Overcurrent protection

• Overcurrent conditions are protected against by both the main fuse and

the separate heating power fuses.

Overheating protection

An error is indicated and all heating is stopped by the microprocessor –

control if the instrument detects any of the following conditions:

• Abnormally high temperature (>75 °C)

• Contradictory results of the temperature sensors

• Failure of one or more heating power control components
• If the microprocessor fails to interrupt heating power, independent tem-

perature limiting hardware circuits limit the temperature rise to a safe

level.

• If the temperature limiting circuits malfunction, an independent hard-

ware thermal fuse circuit cuts power to the heating elements.

Over vacuum protection

• The vacuum system is not capable of generating a dangerous vacuum

condition.

Overcurrent protection

• Overcurrent conditions are protected against by both the main fuse and

the separate heating power fuses.