Operation – Leica Biosystems ASP200 S User Manual

Page 60

Advertising
background image

60

Instructions for Use V 1.4, RevE – 03/2013

Filling with liquid paraffin

Software control

If the instrument has been filled

with liquid paraffin, the instrument's

internal calculation of the melting

time can be disabled.
To do so, in supervisor mode, open the

SERVICE FUNCTIONS (

Fig. 61

) window.

Press PARAFFIN MELTING.

Fig. 62

5. Operation

• If filling with already melted paraffin, the liquid

level must not exceed the MAX level mark (see

Fig. 60, 1

) in the paraffin reservoir.

• After a paraffin reservoir is filled with paraf-

fin, it must be marked as full in the REAGENT

STATUS table.

A safety prompt ap-

pears that you have to

confirm with YES.

In the time window,

enter the desired de-

lay time and confirm

with OK.

Fig. 61

The temperature of filled liquid paraffin must not be greater than 70 °C to prevent damage to the

thermal fuse.

Change the delay time only if you are absolutely sure that the entered time is correct.

If the instrument attempts to run a paraffin step with paraffin that is not completely melted,

substantial faults and malfunctions may result.