Cleaning and maintenance, 2 periodic cleaning and maintenance (continued) – Leica Biosystems ASP200 S User Manual

Page 82

Advertising
background image

82

Instructions for Use V 1.4, RevE – 03/2013

7. Cleaning and Maintenance

Fig. 87

7.2.2 Periodic cleaning and maintenance (continued)

44

Fig. 88

Fig. 89

Use the cleaning tool together with the microfiber cloth (

Fig. 87

) as shown in

Fig. 89

.

45

46

47

The level sensors must be cleaned of all residues (reagents and paraffin) after every

cleaning program!

You will be prompted to clean the sensors by the instrument software at the end of each

cleaning program (

Fig. 88

, 44).

NEVER use the cleaning tool WITHOUT the accompanying microfiber cloth, as other-

wise the sensors will be scratched!

The cleaning tool has two different spoons. The larger spoon (

Fig. 87

, 46) is for use

with the ASP6025, the smaller (

Fig. 87

, 47) is for use with the ASP200 S / ASP300 S.

To prevent damage to the sensors, NEVER use the larger spoon for cleaning the sen-

sors in the ASP200 S / ASP300 S.

Cleaning the level sensors

Advertising