Initial operation – Leica Biosystems ASP200 S User Manual
Page 28
28
Instructions for Use V 1.4, RevE – 03/2013
Fig. 8
Inserting the volume replacement body
26
25
27
25
19
Important!
The instrument must be operated only
when the volume replacement body is
in the retort (even during cleaning pro-
grams).
Otherwise, malfunctions may cause
damage to the instrument.
• Take the volume replacement body (
Fig. 8,
25)
at the metal clip (26) and insert it into the retort
(19) such that the bore (27) is above the posi-
tion of the magnetic stirrer.
4. Initial Operation
4.3 Installing accessories (continued)
Fig. 7
Magnetic stirrer
• Unpack the magnetic stirrer (
Fig. 7,
30) and
insert it into the retort.
• Important!
The magnetic stirrer must be inserted such
that the outer curvature (32) of the two stir-
rer blades (31) faces downwards!
(
Fig. 7
)
Do not insert them the other way around, as
then there will be no effective stirring.
32
31
30