Troubleshooting, 1 general, 2 power failure – Leica Biosystems ASP200 S User Manual

Page 70: Power failure

Advertising
background image

70

Instructions for Use V 1.4, RevE – 03/2013

6. Troubleshooting

6.1 General

For all error states, the instrument takes corrective measures to

protect the specimens, except in such cases where continued

functioning of the instrument would be possible only under hazard-

ous operating conditions.

The Leica ASP200 S is equipped with a cutting-edge fault detection and

troubleshooting system.

When detecting a malfunction, the instrument provides the operator with

detailed error messages and corresponding steps to take via the screen.

• All processes are recorded in detail in the "Run Log."

To inspect the processes, the Run Log file can be opened by pressing

the MENU and RUN LOG symbols.

6.2 Power failure

Power failure during an infiltration program
• If a power failure occurs during an ongoing infiltration program, the

Leica ASP200 S calculates the power failure duration into the total dura-

tion of the step taking place; in other words, the respective step is not

lengthened due to the power failure.

• If the power failure continues for longer than the remaining duration of

the step during which it occurred, the Leica ASP200 S does not shorten

the following steps.

• If, on account of a power failure, the paraffin temperature is reduced to

the point that the paraffin is no longer suitable for continuing the ongo-

ing program, the Leica ASP200 S builds in a delay time sufficient to bring

the paraffin back to a suitable operating temperature before the paraffin

step.

Power failure during a retort or paraffin cleaning program
• If a retort or paraffin cleaning program is running during a power failure,

the total program duration is lengthened by the duration of the power

failure. This is to make sure that the cleaning results are not impaired.