7 switching the instrument on, Switching the instrument on, Initial operation – Leica Biosystems ASP200 S User Manual

Page 32

Advertising
background image

32

Instructions for Use V 1.4, RevE – 03/2013

4.7 Switching the instrument on

4. Initial Operation

• After being switched on, the instrument will take

a few minutes to initialize. See opposite for cor-

responding touchscreen display (

Fig. 14

).

• If the retort is not locked, a signal tone sounds for

approx. 10 sec. during initialization of the instrument.

It is not possible to lock the retort during this time.

• If the retort is locked, no signal tone sounds.

A signal tone sounds for approx. 10 seconds (wait-

ing time) when attempting to open the retort. The

retort cannot be opened during this time.

• After that, the FAVORITES screen will be displayed

(

Fig. 15

).

Screen saver
• A screen saver will turn off the screen display if no

key has been pressed for a (user-programmable)

time. Press any part of the touchscreen to restore

the screen.

After having been restored, the on-screen func-

tions will be nonoperational for a few seconds to

avoid accidental activation of any keys.

Fig. 15

Fig. 14

The instrument MUST be connected to a grounded power socket.

For additional electrical fuse protection, we recommend connecting the Leica ASP200 S to a

socket with a residual current circuit breaker.

Connect the power cable to the power socket. If applicable, switch on the switch for

the power socket.

To switch on the instrument, press in the ON/OFF switch on the rear panel of the

instrument (

Fig. 4,

42) (ON).

Switch on the ON/STOP switch on the right of the instrument (ON).

The ON/OFF switch and the side ON/STOP switch should remain switched on at all times to

keep the heater running for the wax baths.

The ON/STOP switch can be used in emergencies to stop the instrument during while a proce-

dure is in progress.