Operation – Leica Biosystems ASP200 S User Manual

Page 38

Advertising
background image

38

Instructions for Use V 1.4, RevE – 03/2013

PROGRAM OPTIONS
Select the options with which programs are executed here. Press a button

to change the associated value.
STATION SEQUENCE: By Age or Sequential
By Age - RMS is activated, reagents will automatically be used in

order of increasing cleanliness.

Sequential - reagents will be used in sequential order of stations.
WAX BATH ORDER:

Auto or 1. ; 2. ; 3.

Can only be activated if STATION SEQUENCE is set to "Sequential" -

otherwise the order of baths will be managed by the RMS.

PROMPT FOR NUM BLOCKS:

Enabled or Disabled

Enabled- When starting a program, the actual number of blocks must

be entered. This is recommended with the RMS being en-

abled.

Disabled - Number can be entered, but does not have to be.

DATE / TIME
Be sure to verify that the date and time entered do actually correspond

to the local time/date, as this will ensure that all programs are carried out

correctly.
Press the buttons to access the entry windows where settings or correc-

tions can be made, if necessary.

5. Operation

Date and time setup screen

Fig. 27

Fig. 25

Fig. 26