4. Instrument Setup

4.2 Electrical connection

• Check the voltage label (

Fig. 3

) on the rear of

the instrument to ensure that the instrument

delivered is set to the correct voltage range.

Fig. 3

Severe damage may occur if the in-

strument is connected to a power sup-

ply voltage other than that to which it

was originally set.

The power supply voltage for the

instrument is factory preset and

CANNOT be altered by the user.

Notice!

Observe the following instructions

carefully to prevent damage to the

instrument (refer also to

Chap. 2.2

"Warnings – Transport and installa-

tion"

).

41

4.2.1 Connecting the power supply

43

37

38

40

42

Left Fig.

Electrical ports on the rear

panel of the instrument.

Fig. 4

Right Fig.

Correct connection of the cables.

Once the instrument has been switched on, the main switch (ON/OFF) (5) should always remain

in the ON position.

• Connect the power cable (37)

to the input socket (41) of the

main power supply.

• The jumper cable provided

(38) connects the output of the

main power supply (43) to the

input of the electronics module

(40).

Important!

The specification for the

connection (40) is speci-

fied as follows:

100 - 120 V or

230 - 240 V,

maximum 200 VA.