5 anti-reflection clip – function, Anti-reflection clip – function, Initial operation – Leica Biosystems ASP200 S User Manual

Page 30

Advertising
background image

30

Instructions for Use V 1.4, RevE – 03/2013

4. Initial Operation

Changing the clip
• Once the clip is installed, the handle (4) of the speci-

men basket cannot be removed, as otherwise the clip

will be bent.

• To remove or replace a clip, all parts that are inside

the basket (e.g. spiral insert and separating walls)

must be removed from the basket.

• Suitable tool (screwdriver 5,

Fig. 12

).

5

4

Fig. 12

Once bent, clips may no longer be used and

must be disposed of. In this case, install a new

clip from the standard scope of delivery.

The anti-reflection clip (Order no. 14 0476 44135) prevents any reflections that might otherwise

be emitted from the specimen basket (3, order no. 14 0476 34193) and that could affect the level

sensor in the retort, thus causing malfunctions.

Fig. 10

Fig. 11

Installing the clip

1

2b

2a

1

3

4

4

• Insert the left lug (2a) of the anti-reflection clip (1)

into the seventh hole (from the left) of the lower oval

openings (

Fig. 10

).

• Snap the right lug (

Fig. 11

, 2b) into the seventh hole

(from the right) with gentle pressure.

The anti-reflection clips are intended for

installation on the specimen basket to prevent

reflections of the level sensor.

When inserting or removing the metal baskets

into/from the retort, always do so carefully so

that they do not touch or become caught on

the retort wall.

4.5 Anti-reflection clip – Function