7. Cleaning and Maintenance

7.1 Clean programs

The cleaning programs are at the end of the list of all pro-

grams (

Fig. 45

). To display a clean program, highlight the

respective line and press EDIT.

Fig. 81

7.1.1 Retort cleaning programs

A cleaning program comprises a

minimum of three and a maximum

of four steps:

1. Solvent

1 step from station 11
2. Cleaning alcohol

1 step from station 12
3. Cleaning water

1 step from station 13

Optional between steps 2 and 3:

Cleaning water/detergent

1 step from station 13-ext.

Three retort cleaning programs and a paraffin cleaning program are defined in the ASP200 S.

Clean programs cannot be copied or deleted.
Replacement schedule for cleaning reagents:

The cleaning reagents (cleaning xylene and cleaning alcohol) must be replaced weekly (

refer

to Chapter 3.6.1

).

If fatty tissue, bloody tissue, biopsy sponges or the like are processed, both the reagents and

the cleaning reagents must be replaced more frequently.