4 editing the reagent list, Operation – Leica Biosystems ASP200 S User Manual

Page 45

Advertising
background image

45

Leica ASP200 S

5.1.4 Editing the reagent list

Adding new reagents
1.
Enter the reagent name and allocate the new reagent to the reagent

group it belongs to.

2. Determine how many stations (bottles) work with the new reagent.

3. Fill the stations (bottles) according to the reagent list.

Entering reagent names
Takes place via the REAGENTS menu option.
You must be logged on at supervisor access level to proceed.
• From the start screen, press MENU in the MENU FUNCTIONS window;

there, press the MORE button.

• The MORE MENU FUNCTIONS window appears. There, press the

REAGENTS button.

• The SET UP REAGENTS AND WARNING THRESHOLDS window opens.
• To add a reagent:

• Press INSERT to display the keyboard.

• Enter the new reagent name.

• Press OK to confirm.

• You will then automatically be prompted to select the reagent group:

Fig. 40

Select the reagent group

Allocate the new reagent to the desired group

and press OK to confirm.

5. Operation

Fig. 41

Advertising