7. Cleaning and Maintenance

Replacing the active carbon filter

• The life of the active carbon filter will depend

on the reagent types used and the frequency

of vacuum cycles.

• The filter (13) should be replaced at least every

3 months. (

Fig. 86

)

After replacing the filter, open the SMART

MONITOR menu to reset the filter age to "0".

For additional information, refer to

Chap. 5.5

• We recommend a smart clean prior to chang-

ing the filter.

Fig. 86

Draining the condensate container

• Remove, empty, and reinsert the blue conden-

sate container (

Fig. 85

).

Fig. 85

Filter label

14

13

Dispose of waste solvents with care

according to local regulations and the

waste management policy of the com-

pany or institution.

When inserting the new filter, make

sure that it is inserted with the cor-

rect side up, as specified on the label

on the front side of the filter.The arrow

must point upwards.