Leica ASP200 S

5. Operation

Remote draining
• On the screen, select the reagent container

(No. 3) and the receptacle (External).

Then press STATION DRAIN.
• Check that no reagent has leaked.

Remote filling
• On the screen, select an empty reagent

container (No. 3) and the external receptacle

(External).

Then press the STATION REFILL button.
• The remote filling should be completed in less

than 170 seconds.

• Check that no reagent has leaked.

Select station

Press button

Filling/draining reagents
From the FAVORITES screen, press MENU. In

the MENU FUNCTIONS window, tap the SMART

SCREEN button.
The SMART SCREEN (

Fig. 55

) pops up. The smart

screen is the initial screen for manual operation

of the instrument.

Fig. 55

→

Fig. 56

Status bar,

displayed in supervisor mode.

Status bar
The following values are displayed.
• Pressure and temperature of the retort

• Wax bath temperature

• Value of the level sensor of the retort

• Retort lid - closed or open

The following steps should only be

carried out by trained laboratory

personnel, experienced in handling

reagents.