7. Cleaning and Maintenance

Editing steps

• To edit a program step, highlight the corresponding line and press the respective table header.
• To disable a step, set the number of cycles to "0". Only cleaning water steps can be disabled.
The following characteristics of each step can be edited:
Reagent name - only those reagents that are compatible with the reagent station being edited

will be displayed.

Number of cycles - the duration displayed is calculated based on the number of cycles.
Retort temperature - temperatures from 50 to 65 °C can be set.

The finish times of clean programs cannot be edited. Program duration is calculated auto-

matically based on the number of cycles set.

Station 13-ext
• The designation "13-ext" is used for an additional station that can be used in place of the

standard station 13.

• Station no. 13 always contains cleaning water. For some applications it is desirable to use a

cleaning water/agent mixture between the cleaning alcohol and the cleaning water.

• If a program contains the step "Station 13-ext", the operator is prompted to insert the reagent

container "13-ext" at the required point in time during the cleaning program, then to remove

it once it is no longer used.

The station 13-ext can be assigned exclusively to reagents from the cleaning water

group. It may only be used for cleaning programs.

7.1.1 Retort cleaning programs (continued)

• The steps of retort cleaning programs are listed in the order in which they are carried out

(

Fig. 81)

.

• The cleaning program skips any steps not required to complete the cleaning. If, e.g. an infiltra-

tion program has been aborted during the alcohol step, the subsequent clean program will

skip the cleaning solvent step and will start with the cleaning alcohol step.