2 replacing the paraffin, Operation – Leica Biosystems ASP200 S User Manual

Page 58

Advertising
background image

58

Instructions for Use V 1.4, RevE – 03/2013

5. Operation

5.2.2 Replacing the paraffin

Draining the paraffin
• Connect the paraffin drain hose (36) to the

spout (17) of the paraffin drain at the front of

the instrument (

Fig. 57

) and into a catch tank.

Fig. 57

• Select the paraffin reservoir to be emptied and

the receptacle (External,

Fig. 58

).

Then, press STATION DRAIN....

• This is followed by multiple security prompts

that you have to confirm individually in order

to continue.

17

36

From FAVORITES, call up the SMART SCREEN

screen. (

see previous chapter

)

+

Fig. 58

When connecting the hose, be sure to

press it onto the O-rings of the drain

opening as far as it will go.

The paraffin drain hose must be

securely hooked into the external re-

ceptacle and remain there through the

entire drainage process.

When drainage is finished, the hose is

cleaned automatically using air.

Do not remove the hose from the exter-

nal receptacle until this cleaning step

is complete.

Advertising