5.3.1 Editing a program that has been called up.
Fig. 65
To modify a program before the start, press
the EDIT button. A message appears that the
changes will affect running programs only. After
confirming with YES, the program edit window
appears (
Fig. 66
).
Fig. 66
Resetting the end time:
• Touch the FINISH TIME button and reset the
end time in the input window.
Fig. 67
↓
• Enter the day on which you want the program
to end.
• Enter the time (in 24-hour format) at which you
want the program to end.
• Press OK to confirm.
5. Operation
→
↓
Programs can be started up to 6 days in
advance. So that the program is ended
at the defined date/time, the delay step
is lengthened accordingly.