6.4 Typical fill or drain problems

Fill and drain problems may be due to several causes:
1. There is insufficient reagent
• Verify that the fill level of the reagent containers and the paraffin

reservoirs is sufficient.

2. Insufficient pressure or vacuum
The instrument is unable to create sufficient vacuum (for a fill) or pres-

sure (for a drain) in the retort.

• Verify that the reagent container in question is correctly inserted in

the reagent cabinet.

• Verify that all reagent containers are fully pushed home into their

sockets.

• Check whether there is some debris under the retort lid seal caus-

ing a leak.

3. There is a blockage in the air or reagent lines
Blockages in the reagent lines are typically caused by paraffin or tis-

sue debris. If there is reagent in the retort that cannot be drained, try to

remove the blockage as follows:

• Heat the retort to the maximum possible temperature (consistent

with the reagent currently in the retort, if any). Leave the retort at

this temperature for at least 15 minutes.

• While the retort is at maximum temperature, attempt to fill or drain.

• If there is no reagent in the retort:

Run the extended retort cleaning program. Use the maximum set-

tings for number of clean cycles and temperature (consistent with

the type of reagents/solvents used in the clean program).

If you cannot eliminate the problem by carrying out the above list-

ed steps, call Leica Technical Service before attempting to run any

further programs.