2 connecting an uninterruptible power supply (ups), Initial operation – Leica Biosystems ASP200 S User Manual

Page 26

Advertising
background image

26

Instructions for Use V 1.4, RevE – 03/2013

• Connect the ASP to a grounded socket using

the power cable (37).

• Using the jumper cable (38), connect the output

of the main voltage supply to the UPS input.

• Using the jumper cable (38) connect the input

of the electronics section (40) to one of the

UPS outputs.

• Switch the instrument on at the main switch.
• Start the UPS.

4.2.2 Connecting an uninterruptible power supply (UPS)

38

40

37

38a

Fig. 4a

is provided as an example only

and shows the correct connection

of the ASP200 S to an uninterruptible

power supply (UPS).

Fig. 4a

An uninterruptible power supply (UPS)

protects machines and instruments

against malfunctions in the power

supply.

Leica recommends using an

active tracking UPS (with an output

power of 1000-1500 VA) to protect the

instrument and the specimens from tem-

porary power failure, voltage spikes,

undervoltages and excess voltages.

4. Initial Operation

Advertising