7. Cleaning and Maintenance

Step

Daily

Weekly

Monthly

All

3 months

Check the filter screens of the paraffin res-

ervoirs and clean if necessary.

Check the air circulation opening of the par-

affin reservoirs and clean if necessary.

Check and empty the condensate container.

Clean the inlet openings.

Check the status of the active carbon filter

.

Replace the active carbon filter.

Check the level sensors and clean if

necessary.

Check the electrical ports on the rear side of

the instrument.

Clean the reagent containers from the inside

if necessary.

Make sure that the shutter caps of the

reagent containers and the O-rings fit cor-

rectly and are tight.

Check that the reagent containers are fitted

securely in the ports.