133

Displaying the results

The upper part of the results screen shows a correlation plot (green), sensitivity plot (magenta), and
the spectra that correspond to the lowest and the highest constituent values (red and blue). The
lower part of the screen contains a series of graphs that can be displayed by clicking on the
appropriate tab:

45 degree plot (calculated vs. lab data)

residuals vs. calculated values

residuals vs. lab values

calculated % (prediction of the calibration set)

summary of results showing the wavelengths, calibration constants, correlation, standard

error of calibration and F statistics

average differences control chart X with confidence limits

range of differences control chart R with confidence limit

term intercorrelation table

The graphs can be displayed by clicking on the proper tab at the bottom of the display. If the
validation set or outlier set was saved, they can be displayed by clicking on View and selecting the
proper set to be shown.