Displaying the results
The upper part of the results screen shows a correlation plot (green), sensitivity plot (magenta), and
the spectra that correspond to the lowest and the highest constituent values (red and blue). The
lower part of the screen contains a series of graphs that can be displayed by clicking on the
appropriate tab:
45 degree plot (calculated vs. lab data)
•
residuals vs. calculated values
•
residuals vs. lab values
•
calculated % (prediction of the calibration set)
•
summary of results showing the wavelengths, calibration constants, correlation, standard
error of calibration and F statistics
•
average differences control chart X with confidence limits
•
range of differences control chart R with confidence limit
•
term intercorrelation table
The graphs can be displayed by clicking on the proper tab at the bottom of the display. If the
validation set or outlier set was saved, they can be displayed by clicking on View and selecting the
proper set to be shown.