Select the appropriate frequency, and click on Save and then on Close.

If automatic archiving is to be used, the spectra in Routine Analysis cannot be acquired continuously.
There must be a delay between spectra.

If automatic archiving is to be used, the spectra in Routine Analysis cannot be acquired
continuously. There must be a delay between spectra. The archiving scheduler does not
work when analyzing stored data.

Cluster Analysis, Identify, Qualify, Always do Quant, Only do Quant, Reference
Standardized, Master Standardized, Content Uniformity check boxes

Select the desired combination of qualitative analysis options including Cluster Analysis, Identification,
and/or Qualification. Leave all boxes blank if only quantitative analysis is to be performed.

If the "Always Do Quant if Qualification Fails" option is selected, Vision will perform quantitative
predictions even if qualification fails. The "Always Do Quant" option becomes active only after a
calibration is connected to a product.

If the “Only Do Quant if Qualification Fails” option is selected, Vision will perform quantitative
predictions only when qualification fails.

Check Reference Standardized box if your methods are reference standardized. All elements of the
Operation Method (DCM, qualitative, and quantitative methods) should have the same
standardization type.

Check Master Standardized box if your methods are master standardized. All elements of the
Operation Method should have the same standardization type.

If "Content Uniformity Test" is checked, Vision will perform the test automatically. (This option
requires a MultiTab Data Collection Method, and use of the Multi-Tab Analyzer.)