Check the selected set of options.

Instrument must stabilize before data acquisition. When this option is checked, Vision

will not acquire any data before the instrument reaches the proper operating
temperature.

Performance Test must pass before data acquisition. When this option is checked,

Vision will not acquire any data before the Performance Test passes.

Data Collection Method must match instrument configuration. This option is grayed-

out and cannot be changed. Vision will not acquire any data if the DCM does not
exactly match the instrument configuration.

Run Performance Test after instrument linearization. When this option is checked,

Vision will automatically run the Performance Test every time the instrument is re-
linearized. This is not necessary with the XDS.

Use auto-linearization. When this option is checked, Vision will calculate fine tuning

corrections to the linearization constants every time a reference spectrum is collected.
These corrections are not stored in the instrument but applied only to the next
spectrum collected. It is recommended to have this option on at all times.

Reference Standardization. Use this option to standardize the instrument and acquire

standardized spectra on reflectance measurements. This option must be chosen
before creating a DCM. If this option is selected after creation of a DCM, it does not
apply the change to existing DCMs. You must create a new DCM to use Reference