5-4

ADS TRITON+ Manual

Monitor Firing Sensors during Measurement Cycle
without Wireless Communications
110 mA

Monitor Firing Sensors during Measurement Cycle
with Wireless Communications Idle
140 mA

Monitor Firing Sensors during Measurement Cycle
with Wireless Communications Active
(<10
milliseconds)
360 mA

Wireless Modem Communications Idle 110 mA

Wireless Modem during Data Collection (147 blocks
of data over 150 seconds)
130 mA

RS-485 (ExPAC) Modbus Communications (updating
every 2 seconds)
81 mA

USB (ExPAC) Communications Idle 86 mA

USB (ExPAC) Communications during Data
Collection
(147 blocks of data over 50 seconds) 90
mA

Note: Measurement cycle involves firing Peak Combo
Sensor and Surface Combo Sensor and running MLI.

Mounting the XIO or XBUS

Note: These instructions apply only to applications
involving an XIO or XBUS. If the monitor will not
receive external power through an XIO or XBUS, proceed
to the next section, Mounting the Recommended Power
Supply
.

Mount and secure the XIO or XBUS in such a way that takes into
account the specific environment in which it will be located. For
outdoor installation, mount it in such a way that will provide the
best protection against the elements and excessive heat or cold. The
operating temperature range for the XIO/XBUS is -14

to 122

F