10-38 ADS TRITON+ Manual

Communication Problems

The following tables contain general troubleshooting techniques for
wireless communications.

Problem

Monitor does not answer via wireless communication.

Possible

Causes

Signal strength to the modem may be insufficient.

Battery pack may be dead or below minimum voltage
requirement (7.5 volts - internal).

Wireless network failure.

SIM card is not properly provisioned.

Monitor may have lost communication with network and
cannot re-establish communication.

Connection type may be incorrectly configured in the
monitor.

Antenna may be defective.

Monitor may be defective.

Modem may be defective.

3.3-volt fuse on regulator may have blown.

Leading zeros are included before or within the IP
address.