10-46 ADS TRITON+ Manual

Problem

Individual ultrasonic depth readings (SDEPTH_3) from
Surface Combo Sensor/Ultrasonic Depth Sensor are
consistently equivalent to the value obtained from the
following equation: Pipe Height – Physical Offset +
9.99.

Possible

Causes

Sensor may be faulty.

5-volt fuse on regulator may have blown.

Possible

Solutions

Replace the Surface Combo Sensor/Ultrasonic Depth
Sensor.

Replace the 5-volt blue fuse (when replaceable) on the
regulator.

Caution: Do not connect a potentially shorted sensor to
another monitor until confirming that the sensor is
functional (i.e., not shorted) using a voltmeter. A
shorted sensor will blow the 5-volt (blue) replacement
fuse on the regulator of the monitor.

Problem

Sensor reports abnormal temperatures.

Possible

Causes

Sensor connection to monitor may be loose.

Temperature sensor may be faulty.

Sensor may be faulty.

Possible

Solutions

Secure sensor connection to monitor if necessary.

Replace the sensor.