Configuration and Activation 6-9
6. Verify or select the method through which you will
communicate with the monitor from the Connect drop-down
list.
Serial Choose this to perform direct, serial
communication with the monitor.
Note: Always use Serial communications to perform the
initial monitor activation for monitors using the FTP
Operations device and equipped with a SIM with a
dynamic IP address.
Wireless Choose this for TRITON+ monitors with SIMs
with static IP addresses for performing wireless
communication.
FTP Choose this for performing wireless collects,
activates, or firmware updates for monitors using SIMs
with dynamic IP addresses (and configured with the FTP
Operations device) to communicate with the designated
FTP site.
7. (applicable only to Wireless communications) Enter the IP
address corresponding to the monitor in the IP Address field.
The field designation displayed depends on the method of
communication selected in the Connect field.
8. Enter or use the up/down arrows to designate the difference in
hours between your location (or the location of the computer or
network on which the directory resides) and the location of the
monitor in the Time Zone Offset field. For example, if you are
in the Central time zone (in the U.S.) and the monitor is in the
Eastern time zone, enter 1 in the field. However, if you are in
the Central time zone and the monitor is located in the Pacific
time zone, enter -2 in this field.
9. (applicable only to monitors supporting Modbus and Telog Ru-
33 operations) Click on the Modbus button and complete the
Modbus Parameters dialog in the following way:
Select the Enable checkbox.