LABEC M-MIA-LA User Manual
Page 101
 
Ultrasonic milkanalyser
Operation manual
04.11.09
99/107
Cond Measuring
Yes
No OK
Yes
6.Corrections in conductivity measurement
It is done by starting the menu Corrections -> Cond measure. You have the 
possibility to increase/decrease the measured conductivity value from – 1.00 
till +1.00, with step 0.01. After starting this function the display shows the 
following: 
Con Meter
-1.0<=Corr>=1.0
Co Corr=+0.00
Edit – Up/Down
The cursor is positioned below the +. By using buttons Up/Down, the 
operator has the possibility to change the value (number). By pressing the 
button Enter, the operator confirms the chosen value and moves to the next 
position for editing it. After the last position is edited, if the correction value is 
within allowed boundaries, the following is displayed: Co Corr Saved, which 
means, that the correction is entered and saved. On the contrary – it returns 
at the beginning and expects valid correction. 
 
7. Conductivity calibration buffer preparation 
 
In order a standard buffer for conductivity measuring to be prepared follow 
the instruction below: 
1. Take the packet with the powder buffer. 
 
2. Carefully shake the packet in order to gather the powder at the bottom. 
3. Cut one end of the packet.
4. Empty its content in a measuring mug with 1 l volume, paying attention
all its content to be emptied.