Operation manual
04.11.09
44/107
If after recalibration “Save calibration” is pressed the new calibration 
settings will be saved over the factory settings. After that is impossible 
to restore the factory settings of the calibration. Save the newly made 
calibration only if you are sure about its correctness. 
 
7.2.3.7. Settings Page 2.
After this menu is started the display shows the following:
Settings Page2 
Final clean Cnt 
Auto Print Res 
Larg Res En/Dis 
Set Calibr Name 
Exit
Now there is a possibility one of the following options to be set: 
 
7.2.3.7.1. Final Clean Cnt 
Sets the number of cleaning cycles. 
 
7.2.3.7.2. Auto Print Res. 
It is possible (if it is needed) to prohibit the automatic printing the measurement 
results. If there is external keypad connected, then the results could be printed 
by pressing the button 1 from the keypad. It is possible to print out unlimited 
number of printouts for one and the same measurement. 
 
7.2.3.7.3. Larg Res En/Dis. 
The format of the measurement data send towards the computer is set. 
If the option Large Disable, is chosen, then only the main results are send to 
the computer – Fat, SNF, Density, Lac, Proteins, Added Water, sample 
temperature, device serial number and calibration number. 
If the option Enable, is chosen, except the above mentioned parameters also 
data for Salts, Freezing Point, pH, Conductivity will be send to the computer. In 
this case is necessary the software in the computer to be conformable to the 
format of the sent data. After starting the menu, the display shows (for example): 
Large Results
No
No OK Yes