LABEC M-MIA-LA User Manual
Page 37
 
Ultrasonic milkanalyser
Operation manual
04.11.09
35/107
Using the buttons “up” ▲ and “down” ▼ position on the corresponding 
calibration (for example if you need to recalibrate cow milk, which is assigned 
in the milk analyser as Calibration 1 - cow) and press the button Enter.The 
following message appears on the display: 
 
Calibration 1
Getting ready
 
6.4.2.2. Entering values for the separate sample parameters 
Wait for the sound signal and the following menu on the display: 
 
Cal1 Samp High 
Fat=f.ff 
SNF=y.yy 
Den=d.dd 
------------------------------
Lac=l.ll 
Sol=s.ss 
Pro=p.pp 
Exit 
 
 
In this display is possible to be entered the results from the milk with high 
FAT chemical analyses, which were received by the methods from Appendix 
Methods. 
For example: 
- by using the buttons “up” ▲ and ”down” ▼, choose the needed
parameter to be entered;
- by pressing “Enter” the cursor will mark the number to be changed. For
example: FAT=01.29;
- using the buttons “up” ▲and ”down” ▼, enter the needed value; 
- by next pressing of Enter pass towards the next number; 
- after completion of entering the needed FAT value, press Enter; 
- the cursor returns to its starting position. 
 
Cal 1 Samp.High 
Fat=f.ff 
SNF=y.yy 
Den=d.dd