Ultrasonic milkanalyser

Operation manual

04.11.09

42/107

Forwarding the printer could not be necessary if there is a special cable –
fork-joint (from the company-producer), which allows both the keypad and
serial printer to be connected towards connector Printer. In this case the
connector RS232 is free and may be used for communication with computer
IBM PC type.

7.2.3.3. LCD Setup.
Serves to adjust the backlight of the display. The two parameters may be
changed in two modes:

- fast – for rough adjustment;
- fine – for precise adjustment.


After starting this function the displays shows the following:


LCD Setup:
B. Light fast
B.Light fine
Exit


7.2.3.4.Recalibrate.
Serves for changing definite calibration. Methods are described in point 6.4.

7.2.3.5. Edit FrPoint
Through this menu you have the possibility to enter the basic freezing point
separately for each calibration. For more information, see Appendix Freezing
point.

After choosing the menu the following is displayed:


Edit FrPoints
FrPoint Calibr1
FrPoint Calibr2
FrPoint Calibr3
-----------------------------
Exit


After choosing freezing point for the calibration, the following is displayed: