LABEC M-MIA-LA User Manual
Page 38
 
Ultrasonic milkanalyser
Operation manual
04.11.09
36/107
With the button “down” ▼ move the cursor across SNF and analogical to the 
above described consequence enter the value for SNF and after that the rest 
of the values. 
Using menus Edit samp's 1(2, 3) the operator has the possibility preliminary 
to prepare the data for the parameters for each recalibration separately or to 
check the sample data. 
 
6.4.2.3. Making recalibration with the available samples 
After entering the values for the separate parameters of the sample position 
the cursor across Exit, press Enter and the following menu appears on the 
display: 
 
Recalibrate 1
Put sample High 5
times
 
Which reminds us to put 5 times the sample with high FAT. 
 
The sample has to be with temperature in the boundaries 15-25°С. 
 
Before each measurement: 
1.Stir 2-3 times the milk sample before each measurement by pouring it from 
one vessel to another. 
2.Pour the needed quantity in the sample-holder and put it in the recess of 
the device. 
3.Start the measurement. 
4. The sample is sucked. 
Appears the following menu: 
 
Recallibrate 1
Put sample high 5
times
Samp T=16.8
 
After the sample is measured, appears the following menu: