LABEC M-MIA-LA User Manual
Page 86
 
Ultrasonic milkanalyser
Operation manual
04.11.09
84/107
If you’ve entered for the device basic freezing point –0.540
°C, measured
freezing point –0.540
°C, the milk analyser will indicate 0%. When you add
3,8% water, the device will indicate 3,8%-added water. 
From the above mentioned follows that it is very important to enter correct 
basic freezing point in the device. 
 
 
The device’s results for added water may give information about doubt of 
added water in the milk and the exact value of this added water may be 
determined after a “cowshed sample” is taken and the result for the freezing 
point, measured by the milk analyser of the “cowshed sample” is entered as 
basic freezing point in the formula for calculation of added water. 
 
Then the result from this formula will give us the absolute value of the added 
water for the corresponding milk supplier.