29/107
6. MAKING CORRECTIONS AND RECALLIBRATION OF THE 
DEVICE 
At least once monthly check the correctness of the results and need of 
corrections and calibration. 
 
In the process of work with the analyser there is a possibility the results to 
start differing between the data for some of the measuring parameters when 
measured with the milk analyser and the corresponding reference method of 
analyses (Gerber for fat, Kjeldhal for proteins etc). In order to establish the 
possible discrepancy and to correct the readings of the milk analyser do the 
following: 
6.1. Taking samples and preparation of samples for checking the 
accuracy of the milk analyser, making corrections and recalibration 
This is a basic moment for the correct checking the accuracy of the analyser 
and for making correct and precise correction and calibration. It is 
accomplished according Appendix Taking and preparation of samples for 
checking correctness of the milkanalyser, making corrections and 
recalibration. 
6.2. Determination the type of the discrepancy:
6.2.1. Making measurements
Make measurements with different samples (not less than 3) with known 
values of a separate parameter (for example fat content), determined by the 
known reference methods of analyses (for example Gerber's method for 
determination of fat content). For more accuracy it is recommended among 
these samples to be also such with values, close to the lowest and highest 
bounds for the measured parameters. 
Make 5-time measurement for each of the samples. Calculate the average 
value for each sample parameter, without taking into consideration the first 
measurement for each sample. 
6.2.2. Analyzing the measurement results
Make comparison between the values of the parameter from the reference 
sample and measured with the analyser. Make analyses of the difference 
received.