APPENDIX 4 SAMPLING AND PREPARATION OF SAMPLES 
FOR VERIFICATION THE ACCURACY OF THE MILK 
ANALYSER, MAKING CORRECTIONS AND RECALIBRATION. 
 
1. Necessary consumables and devices 
 
- Distilled
water;
- Minimum 3 milk samples with known content of fat, SNF, protein,
density, lactose, solids;
- Heating
water
bath;
- Cooling water bath or chamber;
 
2. General 
 
Milk sampling and storage of samples of raw, thermally treated milk and its 
derivatives (cream, whey, butter-milk etc.) aiming verification the accuracy of 
the analyser, making corrections and recalibration is accomplished following 
the recommendations below: 
- Sample to be taken from homogeneous batches, observing all the
requirement;
- The sample’s volume to be enough for making minimum 5
measurements for each sample or not less than 0.5 l;
- The samples to correspond to the standard physic-chemical and
microbiological requirements, to be pure, without admixtures, without 
added cleaning or other unusual substances and without falsifications; 
- Do not use samples with total acidity of milk more than 17
o
T;
- Vessels, where the samples will be handled have to be clean, dry, glass,
metal or other suitable material, to be tightly closed with rubber or other 
stopples. The stopples not to absorb water and fat and not to influence 
the analyzed sample content; 
- Till the start of the analyses the samples are stored in conditions,
assuring preservation of their content and quantities (advisable low 
temperature – 1-3 
o
С).
For longer storage of the samples a preservative is added as was already 
described in p.9.1.1, and then the sample has to be well stirred. 
 
3. Representative Samples 
 
The samples have to be representative for the corresponding milk type. 
Changes in the analyzed parameters in the samples have, if possible, to 
cover the whole measuring range – i.e. used samples to be with low, middle 
and high content of the analyzed components.