LABEC M-MIA-LA User Manual
Page 89
 
Ultrasonic milkanalyser
Operation manual
04.11.09
87/107
pH electrodes have junction, which allows the internal fill solution of the 
measuring electrode to leak out into the solution being measured. The 
 
junction can become clogged by contamination in the solution. If a clogged 
junction is suspected it is best to clear the junction. 
Heat up the diluted KCl solution to 60-80°C. Place the sensing part of the pH 
electrode into the heated KCl solution for approximately 10 minutes. Allow the 
electrode to cool while immersed in some unheated KCl solution. 
 
- Protein Deposits 
Prepare 1% pepsin solution in 0.1 M HCI. Allow the electrode to stand in this 
solution for five to ten minutes. Rinse the electrode with de-ionized water. 
 
2.9. Electrode activation
Generally, if the procedure of storage and maintenance had been closely 
followed, the electrode can 
be used immediately. However, should the electrode response become 
sluggish, it may be possible that the bulb has dehydrated. 
The bulb can be dehydrated by immersing the electrode in an ideal storage 
solution (e.g. buffer pH 4 solution) for 1-2 hours. If this fails, the electrode 
may require re-activation. If the above procedure does not reactivate the 
electrode to acceptable status, try rejuvenation the electrode by following the 
procedure outlined below. 
 
2.10. Rejuvenation Procedure
Dip and stir the electrode in freon or alcohol for 5 minutes. 
Leave the electrode in tap water for 15 minutes. 
Dip and stir the electrode in concentrated acid (HCI, H
2
S
4
) for 5 minutes.
Leave the electrode in tap water for 15 minutes. 
Dip and stir in strong base (NaOH) for 5 minutes. 
Leave the electrode in tap water for 15 minutes. 
Test with standard calibration solution. 
Finally, test with standard calibration buffer solution to see if the electrode 
yields acceptable results. You may repeat again for better response 
(maximum 3 times). If the response does not improve, then the electrode has 
completed its useful life. Replace with a new electrode. 
 
2.11. Electrode Lifespan
pH electrodes have a finite lifespan due to their inherent properties. How long 
a pH electrode will last will depend on how it is cared and the solution it is 
used to measure. Even if an electrode is not used it still ages. Electrode 
demise can usually be characterized by a sluggish response, erratic readings 
or a reading, which will not change. When this occurs an electrode can no 
longer be calibrated. pH electrodes are fragile and have a limited lifespan.