4.2. Preparation of cleaning solution
4.2.1. Preparation of alkaline cleaning solution 
Preparation of 1 % alkaline solution of Lactodaily for circulation cleaning in 
the milk analyser: 
1. Take the package 100 g concentrated powder chemical Lactodaily 
2. Carefully cut the upper end, paying attention not to spill it. 
3. In appropriate vessel (for example bucket) pour 1 l water. 
4. Add the powder and then again water up to 10 l. 
Then follow the instruction for milk analyser cleaning. 
 
4.2.2. Preparation of acidic cleaning solution 
Preparation of 1 % acidic solution of Lactoweekly for circulation cleaning in 
the milk analyser: 
1. Take the package 100 g concentrated powder chemical Lactoweekly 
2. Carefully cut the upper end, paying attention not to spill it. 
3. In appropriate vessel (for example bucket) pour 1 l water. 
4. Add the powder and then again water up to 10 l. 
Then follow the instruction for milk analyser cleaning. 
 
For final cleaning with the syringe it is recommended to use 1% acidic 
solution of Lactoweekly, preliminary heated up to 70 ˚C. 
 
 
 