LABEC M-MIA-LA User Manual
Page 88
 
Ultrasonic milkanalyser
Operation manual
04.11.09
86/107
- Store the electrode as mentioned above (see section Storage). 
 
2.5. Electrolyte Replacement (for refillable electrode only).
The reference electrolyte needs to be refilled when the electrode has been 
used for a long period, or when the internal electrolyte has dried up. To 
accomplish this, follow the procedure described below. 
- Remove the protective rubber cap or sleeve; 
- Remove the protective rubber sleeve to expose the filling port of the 
electrode;
- Remove the old reference electrolyte with a syringe; 
- Fill the new reference electrolyte. 
2.6. New electrolyte preparation:
- Open the KCl container; 
- Add in de-ionized water until it reaches the level of 20 ml; 
- Close the container and shake it to dissolve the KCl; 
- Add in fresh electrolyte until it reaches the level of the refilling port. The 
reference electrolyte used should be 3M(Mol) KCl;
- Replace the rubber sleeve.
2.7. Re-use the electrode.
- Rinse the liquid junction with de-ionized water.
If these steps fail to restore normal electrode response, you may attempt to
rejuvenate it (See: Electrode Rejuvenation). 
 
2.8. Electrode cleaning
Electrodes which are mechanically intact can often be restored to normal 
performance by one or combination of the following procedures. 
 
- Salt deposits: 
Dissolve the deposit by immersing the electrode in tap water for ten to fifteen 
minutes. Then thoroughly rinse with de-ionized water. Wash the electrode pH 
bulb in a little detergent and water. Rinse electrode tip in with de-ionized 
water. 
 
- Oil/Grease films: 
Wash electrode pH bulb in a little detergent and water. Rinse electrode tip 
with de-ionized water. 
 
- Clogged Reference Junction: