LABEC M-MIA-LA User Manual
Page 39
 
Ultrasonic milkanalyser
Operation manual
04.11.09
37/107
Recalibrate 1
Put sample High
Cal meas=1/5
 
Follow the procedure till the 5th measurement. 
After 5th measurement completion automatically appears the menu: 
 
 
Cal 1 Samp Low 
Fat = f.ff 
SNF=y.yy 
Den=d.dd 
------------------------------
Lac=l.ll 
Sol=s.ss 
Pro=p.pp 
Exit 
 
 
Which reminds to enter the values for the sample with low FAT? Enter the 
values analogically to the procedure, described in the previous sample. After 
entering the last parameter position the cursor on “Exit”, press “Enter” and 
the following menu will appear: 
 
Recal 1
Put sample Low 5
times
 
Make 5 times measurement of the low FAT sample. 
After 5th measurement completion automatically appears the menu: 
 
Recal 1
Put sample water
 
Which reminds for 5-times water measurement? There is no need to enter 
values after placing the sample-holder. Measurement starts directly.