LABEC M-MIA-LA User Manual
Page 56
 
Ultrasonic milkanalyser
Operation manual
04.11.09
54/107
 
By which the device passes in a mode of measuring high fat samples. 
This calibration mode is active till the power supply of the device is switched 
off i.e. it has to be always set if the device will be calibrated for high fat 
measurement. 
 
8.7. Working with embedded accumulator (option). 
On customers’ request accumulator could be embedded in the milk analyser. 
In this way it could work independently on the electrical network. In this case 
the Power switch on the rear panel of the analyser serves for close / open the 
electric circuit of the outer power supply. After closing the circuit, in order to 
start the analyser, the operator has to press the Start button on the front 
panel of the analyser. 
 
Switching off the analyser could be done on one of the following ways:
Through the analyser – by switching the Power button on the rear panel of 
the analyser. 
Through the software – by using the option Shut Down from the Calibration 
and Working mode menus. 
After pressing and holding pressed the button Enter (when choosing 
calibration), the following is displayed: 
 
 
Milk selector 
Cal1 – Cow 
Cal2 – Sheep 
Cal3 – UHT 
-----------------------------
Cleaning 
Final clean 
Shut Down 
 
 
By choosing the option Shut Down the operator may switch off the power 
supply of the analysers. 
In case that the external power supply from the electrical network, the 
analyzer automatically starts working with power supply from the battery. 
Possible working modes with battery: 
 
1. The analyser is powered by the battery, works normally, the display shows 
the following message 
 
 
Battery Powered