LABEC M-MIA-LA User Manual
Page 95
 
Ultrasonic milkanalyser
Operation manual
04.11.09
93/107
pH Calibration 
pH Measuring 
pH En/Disable 
pH U Display 
------------------------- 
pH Test 
Co Meter Calibr 
Co Meter Test 
Co Meter En/Dis 
------------------------- 
Exit 
8. pH Calibration. 
Serves for calibrating the pH meter. For that purpose use 2 standard buffers, 
shown on the display as Low buffer (for example 5.00 pH) and High buffer 
(for example 7.00 pH). The procedure is following: 
 
1. Start menu Calibration. 
2. Put the pH probe in Low buffer. 
3. By pressing the buttons of the analyser enter the exact buffer value. The 
display shows: 
 
pH Calibr
Put Izopot buff
Buf=xx.xxx
4. Input the buffer value when the probe is in its isopotential point and with 
which our probes are accompanied with. 
Then the display shows: 
 
pH Calibr
Put Izopot buff
Buf=xx.xxx
V=x.xxxV Set
Where x.xxx is measured by the probe voltage.