LABEC M-MIA-LA User Manual

Page 95

Advertising
background image

Ultrasonic milkanalyser

Operation manual

04.11.09

93/107

pH Calibration
pH Measuring
pH En/Disable
pH U Display
-------------------------
pH Test
Co Meter Calibr
Co Meter Test
Co Meter En/Dis
-------------------------
Exit

8. pH Calibration.
Serves for calibrating the pH meter. For that purpose use 2 standard buffers,
shown on the display as Low buffer (for example 5.00 pH) and High buffer
(for example 7.00 pH). The procedure is following:

1. Start menu Calibration.
2.
Put the pH probe in Low buffer.
3. By pressing the buttons of the analyser enter the exact buffer value. The
display shows:

pH Calibr

Put Izopot buff

Buf=xx.xxx

4. Input the buffer value when the probe is in its isopotential point and with
which our probes are accompanied with.
Then the display shows:

pH Calibr

Put Izopot buff

Buf=xx.xxx

V=x.xxxV Set

Where x.xxx is measured by the probe voltage.