Ultrasonic milkanalyser

Operation manual

04.11.09

65/107

5.3. Basic principles.

Milk is dried, carbonized and turned to ashes till constant mass. The ashes
received are calculated in percentages.

5.4. Necessary devices and reagents

- Assay

balance;

- Crucibles;
- Water-bath or infrared lamp;
- Hot plate or burner;
- Drying-oven with thermal regulator;
- Muffle

furnace;

- Exicator;
- Quantity

filter.

5.5. Making the determination:

In preliminary tempered and weighted crucible of the assay balance at about
10 g milk is weighted with accuracy up to 0,0005 g. The crucible with the
sample is placed in a water-bath or infrared lamp till the evaporation of milk to
dry state. Then it is carbonized with the burner or on a hot plate, paying
attention not to be splashed out. The crucible is placed in a muffle oven and
turns to ashes slowly, without the sample to be kindled, at temperature 500-
550

о

С till white or grey-white ashes. It is tempered in an exicator and is

weighted till the appointed accuracy. Heating up in the oven is repeated till a
constant mass is received.

5.6. Calculations

Ashes content is calculated using the formula


Where:
А – the mass of empty, tempered crucible, g
В – the mass of the crucible together with the milk, g
С – the mass of the crucible with the received ashes, g

5.7. Measurement accuracy

The difference between tow parallel determinations could not be more than
0,02 %.

100

)

(

)

(

=

A

B

A

C

ashes