LABEC M-MIA-LA User Manual
Page 67
 
Ultrasonic milkanalyser
Operation manual
04.11.09
65/107
5.3. Basic principles.
Milk is dried, carbonized and turned to ashes till constant mass. The ashes 
received are calculated in percentages. 
 
5.4. Necessary devices and reagents
- Assay
balance;
- Crucibles; 
- Water-bath or infrared lamp; 
- Hot plate or burner; 
- Drying-oven with thermal regulator; 
- Muffle 
furnace;
- Exicator; 
- Quantity 
filter.
5.5. Making the determination:
In preliminary tempered and weighted crucible of the assay balance at about 
10 g milk is weighted with accuracy up to 0,0005 g. The crucible with the 
sample is placed in a water-bath or infrared lamp till the evaporation of milk to 
dry state. Then it is carbonized with the burner or on a hot plate, paying 
attention not to be splashed out. The crucible is placed in a muffle oven and 
turns to ashes slowly, without the sample to be kindled, at temperature 500-
550 
о
С till white or grey-white ashes. It is tempered in an exicator and is
weighted till the appointed accuracy. Heating up in the oven is repeated till a 
constant mass is received. 
 
5.6. Calculations
Ashes content is calculated using the formula
 
Where: 
А – the mass of empty, tempered crucible, g 
В – the mass of the crucible together with the milk, g 
С – the mass of the crucible with the received ashes, g 
 
5.7. Measurement accuracy
The difference between tow parallel determinations could not be more than 
0,02 %. 
100
)
(
)
(
∗
−
−
=
A
B
A
C
ashes