23/107


1. Choose Final Clean. After choosing the final cleaning, the following
message appears on the display:

Put filled with

Water glass

and press Enter

to Continue


2. Put a glass filled with water and press the button Enter in order to continue
the procedure. Then the following message appears on the display:

Cleaning

Please wait

Do not miss to clean with water because the residues from the alkaline
cleaning solution could react with the acidic solution, which will lead to gas
and sediment formation.

After rinsing with water, the display shows the following:

Put filled with

Detergent glass

and press Enter

to Continue


3. Pour warm alkaline solution in the glass and press the button Enter, in
order to continue the cleaning.
Then the following message appears on the display:

Cleaning

Please wait