LABEC M-MIA-LA User Manual
Page 34
 
Ultrasonic milkanalyser
Operation manual
04.11.09
32/107
By using buttons “up”▲ and”down”▼position on the corresponding 
calibration (for example Correction 1 – cow) and press Enter. 
6.3.3.2. Choosing correction parameter 
After choosing calibration mode the display shows the following: 
 
 
 
 
 
 
 
 
 
 
 
 
Using the buttons “up”▲ and”down”▼position on the parameter to be 
corrected (for example FAT) and press the button Enter. 
6.3.3.3. Making correction 
After choosing parameter (for example fat) the display shows the following: 
 
FAT Correction
0,05
- OK
+
 
Using the buttons “up”▲ and”down”▼is possible to increase or decrease the 
value of the measured parameter in the above pointed limits. Leaving this 
mode means saving the correction value and activating it. 
6.3.3.4. Making verification 
After the corrections are made put the milk analyser in working mode 
following the sequence below: 
1. Switch off the power supply. 
2. Switch on the “POWER” button, which starts the identification 
procedure. For a short time the display shows the number of the 
software versions, for example: 
Cal1 Correct’s 
FAT 
SNF 
Density 
------------------------------ 
Lactose 
Salts 
Proteins 
Water 
------------------------------ 
Exit