LABEC M-MIA-LA User Manual
Page 35
 
Ultrasonic milkanalyser
Operation manual
04.11.09
33/107
 
Milk analyser xxx 
LCD vers xx 
MA vers yy 
MA ser. N. xxxx 
 
3. Till the analyser is prepared for work (at about 5 minutes) the following
message is written on the display: “Getting ready”.
4. When the device is ready for work, a beep is heard and the display
shows: “Analyser ready”. The analyser is ready to make analyses in 
mode 1 (normally Cow). 
5. make several times measurement of reference samples with known
values of the corrected parameter. If the difference between the values 
of the parameter from the reference methods and milkanalyser are in 
the limits for the parameter it may be considered that the correction is 
successfully made. 
If the discrepancy between the measurements from the milk analyser and 
classical methods is bigger than is necessary to make second correction 
according above described way. 
If after the second correction the results are unsatisfactory we recommend 
making a calibration of the analyser. In dependence of the conditions and 
your requirements you may make the calibration using a personal computer 
type IBM PC and the company's calibration program or autonomous - by 
recalibration. 
When making corrections or calibrations be 100% sure in the accuracy of the 
reference methods result. 
 
6.4. Recalibrating the milk analyser 
6.4.1. Running the analyser in mode Recalibrate
6.4.1.1. Press the button Enter and without releasing it switch on the power 
supply of the device, wait for the starting identification messages and release 
the button after the following message appears on the display: 
 
Release button
to start setup