Introduction and Specifications

Instrument

Specifications

1

1-9

Instrument Specifications

Modes of Operation

Line Powered (Battery Charge and Maintenance Charge)
Battery Operation
Offline (Battery Charge and Maintenance Charge)

Displayed Parameters

Power (Watts)
HF Current (milliamperes)
Test Load (ohms)

Test Performed

Generator Output
HF Leakage ........................................................ Performs HF leakage tests to IEC 601

2-2, 1289-2, ANSI/AAMI standards:
Type BF Test 1 – Earth referenced
monopolar output Type CF/Bipolar –
Isolated monopolar or bipolar output

Measurement

Technique ........................................................... Precision high-voltage capacitive

attenuator samples applied ESU
signal. This directly measured HF
voltage and the selected test load
resistance value utilized to derive the
true-rms values of both current and
wattage readings.

HF Power (watts)

Resolution....................................................... 1 to 400 W / Resolution: 0.1 W
Maximum power input .................................... 400 W rms

Accuracy..................................................... ± 5 % of reading or ± 3 watts,

whichever is greater

HF Current

Range ......................................................... 30 to 2500 mA rms, Resolution: 1 mA
Accuracy..................................................... ± 2.5 % of reading or ± 15 mA,

whichever is greater.

Bandwidth/System Response

Bandwidth of rms converter circuit (1 % accuracy)

Flat response .................................................. 10 kHz to 10 MHz
-3 dB points..................................................... 1 kHz to 20 MHz

System Response (measurement circuitry and selected test load):

-3 dB points..................................................... 1 kHz to 10 MHz @ 300 e