RF303
Operators Manual
3-12
W
Caution
To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer’s duty cycle, which is 30
seconds ON during any continuous one-minute period.
WHILE IN BATTERY OPERATION MODE:
Use Lead from side battery ground banana
jack to earth reference point.
CQM
DISPERSIVE
ACTIVE
ESU
RF303
CONNECT ONE HF LEAD ONLY
OPEN CIRCUIT TEST
HF Leakage Test 1
OR
HF
METER
YEL
BLU
GRN
M
TYPE CF
F
GROUND EARTH
REFERENCE
Jumper
ESU
Dispersive
ESU
Active
SUPPLY
MAINS
EARTH
MAIN
TEST
LOAD
exz11.eps
Figure 3-3. HF Leakage Test Number One