Fluke Biomedical RF303 User Manual

Page 61

Advertising
background image

Operation

Remote Operation via the RS-232 Function

3

3-25

Table 3-2. Available Duplex Mode Commands (cont.)

Command Description

Parameters

Required

Return Value

(Response)

RDCURRENT Prompts

an

Analyzer

current measurement in

milliamperes. Not

available in RARF

mode.

None AAAA

(mA)

Where AAAA is

current in

milliamperes.

RDMETER

Prompts the Analyzer to

measure the applied

signal of the ESU.

None

Prompts the Analyzer to

return all measurements

of the applied signal of

the ESU.

{MODE=NONE}

None I=AAAA(mA),

CF=XX.X,

KVPP=XXX,

PWR=PPP.P(w)

Where AAAA is the

Current reading in

milliamps, XX.X is the

Crest Factor, XXX is

the peak-to-peak

voltage in kVolts and

PPP.P is the power in

watts.