Fluke Biomedical RF303 User Manual

Page 58

Advertising
background image

RF303

Operators Manual

3-22

All responses are uppercase, followed by CR and LF.

Measurement readings from the Analyzer are returned identically as

they would appear on the Analyzer's display; that is, as numbers, with
or without a decimal point, and followed by any appropriate unit of
measurement.

Errors always return a response beginning with: ?ERR=XX (plus the

description). Table 3-1 lists each of the error responses.

Table 3-1. Error Responses

Error Response

Explanation

?ERR=01,BAD OR MISSING
PARAMETERS

The received command requires
parameters, and there were none, or
they were not recognized.

?ERR=02,ILLEGAL NUMBER OF
PARAMETERS

The command did not include the
correct number of parameters.

?ERR=03,PARAMETER OUT OF
RANGE OR SYNTAX ERROR

One or more of the parameters were
out of the correct range, or there was
an error in the syntax of the
command.

?ERR=04,COMMAND NOT
AVAILABLE FROM THIS MODE

The Model Analyzer is not in the
special sub mode within the remote
that is required for the command.

?ERR=05, DEACTIVATE ESU
BEFORE ADJUSTING

There is an ESU signal present at the
meter inputs that is not allowed for
the command to be executed.

?ERR=06,ACCESS DENIED

The command is a calibration
command and the calibration jumper
is not connected

Table 3-2 describes various commands available in Duplex Mode.